Description
| Channel Features | Constant Current Source and Constant Voltage Source with
Independent pairs of Closed-loop Structure |
|
|
Voltage |
CV Output Range | 25mv-5V |
| Min Discharge
Range |
1.0V for universal holder, 1.5V for 2m cable | |
| Accuracy | ± 0.05% of FS | |
| Stability | ± 0.05% of FS | |
|
Current |
Per Channel Current Range | Range 1: 0.5mA-100mA Range 2: 100mA~3A
Range 3: 3A-6A |
| Accuracy | ± 0.05% of FS | |
| CV cut-off current | Range 1: 0.2mA Range 2: 6mA Range 3: 12mA | |
| Stability | ± 0.05% of FS | |
|
Power |
Per Channel Output
Power |
30W |
| Stability | ± 0.1% of FS | |
|
Time |
Current Response
Time |
<1ms (10% FS~90% FS) |
| Working step time | ≤ (365*24) hour/step, Time Format 00:00:00(h: min: s, MS) | |
|
Data Record |
Data record condition |
Min data record interval: 100mS |
| Min voltage change:10mV | ||
| Min current change: Range 1: 0.2mA, Range 2: 6mA, Range 3: 12mA | ||
| Frequency | 10Hz | |
|
Charge |
Charge Mode | CC, CV, CCCV, CP |
| End Condition | Voltage, Current, Relative Time, Capacity, -△V | |
|
Discharge |
Discharge Mode | CC, CV, CCCV, CP, CR |
| End Condition | Voltage, Current, Relative Time, Capacity, -△V | |
|
Pulse |
Charge | CC, CP |
| Discharge | CC, CP | |
| Min Pulse Width | 500ms | |
| Items | Values | |
| Pulse counts | Up to 32 | |
| Automated Switch | Automated switch from charge to discharge for each pulse | |
| Cut-off Condition | Voltage, Time | |
| DCIR | Calculated DCIR values available in BTSDA | |
| Cycle | Max cycles | 65535 |
| Max Steps | 254 | |
| Cycle nest | 3 | |





Reviews