Semco SI CT 5V 300A 4CH – High-Current Lab Cell Tester for Advanced R&D
High-precision 5 V / 300 A / 4-channel cell tester for lab-grade cycle life, capacity, DCIR and pulse testing. Each channel delivers 1500 W with AD/DA 16-bit resolution, ±0.1% FS accuracy, fast 20 ms current response, and 100 ms data logging. Ideal for high-current lithium cell development, abuse testing and performance characterization with TCP/IP control, MySQL database and rich data export.
Semco SI CT 5V 300A 4CH – 1500 W Per Channel High-Current Cell Testing System
The Semco SI CT 5V 300A 4CH – Neware-2020 (LAB) CT 4004 is a high-end storage cell charge–discharge and pulse testing platform built for demanding R&D and lab environments. Working from AC 380 V ±10%, 50 Hz, it provides 4 channels, each delivering up to 1500 W with 1.5–300 A current capability and 0.025–5 V CV range, making it suitable for high-rate cylindrical, pouch and prismatic cell testing.
With 16-bit AD/DA resolution, ±0.1% FS accuracy on both voltage and current, and a minimum data record interval of 100 ms, the system supports precise CC, CV, CCCV, CP charge and CC, CP, CR discharge. Pulse testing is supported down to 500 ms pulse width, with up to 32 charge–discharge pulses per sequence and full cut-off control by voltage, current, Δtime, capacity and ΔV. DCIR measurement is integrated, and channels can be paralleled for higher current (pulse disabled in parallel).
Key Technical Summary
| Parameter |
Specification |
| Model |
Semco SI CT 5V 300A 4CH (Neware-2020 LAB) |
| Channels |
4, independent control |
| CV output range |
0.025–5 V, Min discharge 2.5 V |
| Current range |
1.5–300 A |
| Output power per channel |
1500 W |
| Accuracy (V / I) |
±0.1% FS |
| Min record interval |
100 ms (10 Hz) |
| Max cycles / steps / nest |
65,535 / 254 / 3 |
| DCIR & pulse |
DCIR supported, 500 ms min pulse |
| Communication & DB |
TCP/IP, Ethernet, MySQL, XLS/PDF/TXT export |
| Aux channels |
Up to 248 temp + 248 voltage per main channel |
The system uses Kelvin 4-wire connection for accurate sampling and offers software protection (power-off data retention, offline mode, user-defined protection) plus hardware foolproof protection (IP20). Operating in 0–40 °C, ≤70% RH, the 30U 600×600×1520 mm cabinet integrates auxiliary temperature (-25–110 °C, ±1 °C) and voltage (-5–5 V, ±0.1% FS) channels with configurable upper/lower limits and ΔV alarms, enabling detailed cell behavior analysis and safety monitoring for advanced battery research.